氧氮氢分析仪的使用注意事项:
1 、氧氮氢分析仪在初次启用前,应该对连接点、焊点、阀门等进行检漏,以确保空气中的氧不会反渗进入管道及仪器内部,造成测量数值偏高。
2、再次使用仪器前,要进行管道系统净化,将漏入的空气除干净,同时确保连接取样管路时没有漏入空气。
3、样气中氧含量的变化会受管道材质及表面粗糙度影响,因此一般连接管路选用铜管或抛光过的不锈钢管,而不用塑料管、橡胶管等。
4、微量元素分析时,要避免各种管件、阀门、表头等死角对样气造成污染,因此必须尽可能的简化气路系统,连接件死角要小,以防止溶解氧逸出造成污染,使用水封、油封及腊封等设备,才能较好的确保数据的。
改进的红外检测池技术。新疆美国分析仪价格
Sample Preparation
Sampling and sample preparation of refractory metals such as titanium and zirconium is somewhat different from that of steel. Unlike steel samples, hydrogen is not as mobile in this group of materials; therefore, storage in liquid nitrogen or dry ice is not required. However, it is important to keep the sample cool when cutting or sectioning. Sample preparation for oxygen and nitrogen determination has beendifferent from that of hydrogen determination.
美国LECO分析仪价格双直流风扇保证持续冷却,不受电压波动的影响。
氧氮氢分析仪的操作过程
氧氮氢分析仪可根据的用户的要求定制。该分析仪配备两个不同灵敏度的探测器,在更多的应用中保证了准确度。
操作简单而且安全。在样品称重以后,样品的重量数据通过电子界面传导至相连的电脑中,也可以在软件界面中手动输入样品的重量。
样品被放置在水平放置的炉子中的低温区域。分析开始以后,炉子开始向上转动,然后样品逐渐落入加热区。通过添加载体气体氮气,氢气被扩散出来并且随载体被传导至一个高灵敏度的热传导测量池。
通常一次测量的时间在3到15分钟之间。在分析的过程中设备可以显示测试信号和设备的参数,从信号的评估到结果在界面上的显示全部自动完成;而测量数据则被传导至LIMS(实验室信息管理系统)中。无需过多的保养和维护,而除尘器和净化管的维护工作也方便。
美国LECO力可氧氮氢分析仪ONH836(氧氮氢联测仪)系列硬件部分
特点和优势
1、炉头区域LED照明系统
2、炉头快速拆装的粉尘过滤装置
3、装有自动旁路阀的可快速拆装氧水捕集装置
4、**的动态流量补偿(DFC)保证热导池气体流量、压力恒定
5、改进的红外检测池技术
新红外池的恒温绝热技术消除了外界环境温度的干扰
长寿命的红外发射光源和无漂移的电路设计保证红外检测池的长期稳定
氧高低双量程采用3个碳红外检测器,同时检测一氧化碳和二氧化碳,保证准确定量
6、高效脉冲炉冷却系统
强力内循环水冷装置
双直流风扇保证持续冷却,不受电压波动的影响
新电极炉头设计,提高脉冲炉热交换效率
7、全自动分析系统(可选)
自动清扫装置
20位或100位全自动分析系统
高速全自动真空除尘装置直接将粉尘排至仪器外部捕集器中,消除环境污染
可选六轴可旋转触摸屏幕,增强了操作的工效和直观性。
Procedure - Solid Samples:
--Instrument calibration/drift correction.
Place the nickel basket containing the calibration/drift standard into the open port at the top of the loading head.
Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.
Clean the upper and lower electrode manually, or, if applicable, remove the crucible and press the analyze button to clean with the automatic cleaner.
Cornerstone氢氧氮分析软件同时支持***的书记文档格式和可变的报告格式。新疆美国分析仪价格
Procedure - Solid Samples
Instrument calibration/drift correction.
Place the calibration/drift standard in a *** Nickel Basket, and if applicable, place the sample into the appropriate autoloader position.
Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open.
Note: samples using automation should be placed in the appropriate autoloader position before starting the
analysis sequence. Once the sequence has started, the automatic analysis will start and end automatically.
SamplePreparationSamplingandsamplepreparationofref
【详情】美国力可LECO氧氮氢分析仪公司介绍:美国力可公司(LECO)始于1936年,以生产用于冶金、机械、...
【详情】Oxygen,Nitrogen,andHydrogeninRefractoryMetalsSumma
【详情】氧氮氢分析仪的优点采用惰气熔融技术,同时测定氧、氮和氢元素含量新设计封闭气路系统,优化气路循环,使氧...
【详情】Oxygen,Nitrogen,andHydrogeninRefractoryMetalsSumma
【详情】Procedure–Powder/ChipSamplesInstrumentcalibration/...
【详情】力可OHN736&836系列仪器采用惰性熔融法,提供各种无机材料、钢铁、有色金属、硬质合金陶瓷材料中...
【详情】SamplePreparationTypically,titaniumandzirconiumsam
【详情】ONH836系列:氧氮氢分析仪(熔融法)软件特点和优势Cornerstone氧氮氢分析软件分成4个主...
【详情】Procedure-SolidSamplesInstrumentcalibration/driftc
【详情】